Model
|
900B (Standard Model)
|
900B (Advanced Model)
|
Display
|
10.4 inch high visibility color TFT display
|
Ventilation Mode
|
A/C, SIPPV, IPPV, SIMV, PSIMV, PSV, CPAP, PCV, VCV, SPONT, MANUAL
|
Minute Volume
|
Greater than or equal to 18L
|
Back-up Power Supply
|
At least 4 hours
|
Tidal Volume
|
Adjustable range: 10-1500ml, Display range: 0-2000ml
|
Respiratory Rate
|
1-99bpm
|
Inspiratory/expiratory (I:E) ratio
|
4:1-1:4
|
PEEP
|
0-10cmH
2
O
|
SIMV Rate
|
1-20bpm
|
Inspiratory Pressure Trigger
|
.
-10-10cmH
2
O
|
Trigger Sensitivity (flow)
|
1-10L/m
|
Inspiratory Plateau
|
Adjustable from 0-50% of inspiratory time
|
Adjustable Oxygen Concentration
|
45-100%
|
21-100%
|
Gas Supply Requirement
|
280
~
600 kPa medical grade oxygen and compressed air supply
|
Pressure Limit
|
1.0kPa~6.0kPa
|
Maximum Safe Operating Pressure
|
≤
6.0kPa
|
Tidal Volume Alarm
|
Upper Limit: 10~2000ml, Lower Limit: 0~1800ml
|
Airway Pressure Alarm
|
Upper Limit: 0.1~6kPa, Lower Limit: 0~5kPa
|
Oxygen Concentration Alarm
|
Upper Limit: 22~100%, Lower Limit: 21~80%
|