Technique Specifications | |
Physical specifications | |
Dimensions: | 52*73*123 cm |
Weight: | 60kg |
Casters: | Front wheels with lock |
Screen: | LCD |
Display: | 7 inch |
Applications : | |
Patient range : | Aud l t / Peadiatric |
Language : | ENGLISH |
Pipeline supply | |
Gas configuration | O 2, N 2 O |
Pipeline input range | 280-600kPa |
Working specifications | |
Power: | AC 220 v ± 10%, 50 H z ± 2 % |
Current: | 500mAax |
Battery Power | |
Battery Type | Lead-acid 12V, 4Ah |
Run-time | Typical 30 mins+, up to 2 hrs |
Modes of ventilation |
IPPV | Intermittent Positive Pressure Ventilation |
SIMV | Synchronized Intermittent Mandatory Ventilation |
MANUAL | Manual |
SIPPV(A/C) | Synchronized Intermittent Positive Pressure Ventilation |
STAND-BY | STAND-BY |
Ventilator parameter | |
Working type | Elect r onical l y control l ed, pneumatic driven |
Tidal Volume range | 40-1500ml |
Pressure range | 5 - 60 cmH 2 O |
Rispriation Rate Range | 1-99b pm (1-40bpm under SIMV ) |
I:E Range | 4:1-1:8 |
Ptrigger Range | -20-0cmH 2 O |
Minute volume range | > 18 L/ Min |
SIGH range | 0-5/100 |
System Standard | |
Hypoxic guard system | N 2 O cut-off valve, O 2 c o ncentra t ion >25% |
Safety valve | <12.5kPa |
O 2 flush | 25-75L/min |
Volume of CO 2 absorber | 2L |
Flow meter | 4 t ubes, Cas c ade 0.1-15 L/ min O 2 |
0.1-15L/min N 2 O | |
Vaporizer | 1 (with optional for Selectetac), maximum 2 mounts( I soflurane/Enflurane/Sevoflurane/Halothane) |
Monitoring | |
Types | VT. MV. BPM. Ppeak, Pm e an, Inspiratory Platform, F i O 2 (Optional) |
Waveforms | Waveforms of P - T, F-T |
Alarm Limits | |
Audio/Visual |
VT too high/too low,MV too high/too low,airway pressure, FiO
2
(Optional)
Oxygen failure,AC power failure,Battery low |
Optional | |
2.Yoke | |
3.AGSS | |
4.Suction Unit |